492 Chapter 3
Performance Verification Test Records
Agilent E4403B Performance Verification Test Record
36. Displayed Average
Noise Level
1 kHz RBW
10 MHz to 1 GHz (1)________ –117 dBm
±1.23 dB
1 GHz to 2 GHz (2)________ –116 dBm ±1.23 dB
2 GHz to 3 GHz (3)________ –114 dBm
±1.23 dB
100 Hz RBW
10 MHz to 1 GHz (7)________ –125 dBm
±1.23 dB
1 GHz to 2 GHz (8)________ –124 dBm
±1.23 dB
2 GHz to 3 GHz (9)________ –122 dBm
±1.23 dB
39. Residual Responses
150 kHz to 3.0 GHz (1)________ –90 dBm
±0.87 dB
42. Tracking Generator
Absolute Amplitude
and Vernier Accuracy
(Option 1DN only)
Absolute Amplitude
Accuracy
–0.75 dB (1)________ 0.75 dB
±0.096 dB
Absolute Vernier
Accuracy
–2 dB –0.5 dB (2)________ 0.5 dB
±0.008 dB
–3 dB –0.5 dB (3)________ 0.5 dB ±0.008 dB
–5 dB –0.5 dB (4)________ 0.5 dB
±0.008 dB
–6 dB –0.5 dB (5)________ 0.5 dB
±0.008 dB
–7 dB –0.5 dB (6)________ 0.5 dB ±0.008 dB
–8 dB –0.5 dB (7)________ 0.5 dB
±0.008 dB
–9 dB –0.5 dB (8)________ 0.5 dB
±0.008 dB
–10 dB 0.5 dB (9)________ 0.5 dB ±0.008 dB
Table 3-6 Agilent E4403B Performance Verification Test Record
Agilent Technologies
Model E4403B Report No. ___________
Serial No. ___________ Date ___________
Test Description Minimum
Results
Measured
Maximum
Measurement
Uncertainty
Przeglądanie stron 491