Agilent Technologies E1401-90493 Instrukcja Użytkownika Strona 171

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 642
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 170
Chapter 2 171
Performance Verification Tests
22. Frequency Response: Agilent E4401B and E4411B
4. Enter the most negative number from the Flatness Relative to
50 MHz column of Table 2-41:
____________ dB
5. Enter the most negative number from the Response Relative to
50 MHz column of Table 2-42:
____________ dB
75
inputs, Option 1DP: The frequency range below 100 kHz was not
tested; no entry from Table 2-42 is necessary.
6. Record the more negative of numbers from step 4 and step 5 in
Table 2-43 as the Minimum Response for Band 0.
7. Subtract the Minimum Response for Band 0 from the Maximum
Response for Band 0 and record the result as the Peak-to-Peak
Response for Band 0 in Table 2-43.
Table 2-43 Frequency Response Results
Band
Maximum Response Minimum Response Peak-to-Peak Response
dB
Test
Record
Entry
dB
Test
Record
Entry
dB
Test
Record
Entry
01) 2) 3)
Przeglądanie stron 170
1 2 ... 166 167 168 169 170 171 172 173 174 175 176 ... 641 642

Komentarze do niniejszej Instrukcji

Brak uwag