Chapter 2 369
Performance Verification Tests
47. Tracking Generator Non-Harmonic Spurious Outputs: Agilent E4401B and E4411B (Option
1DN or 1DQ)
Determining the Highest Non-harmonic Spurious
Response
1. In Table 2-102, locate the most positive non-harmonic response
amplitude. Record this amplitude as the highest non-harmonic
response amplitude in Test Record entry 1 of the performance
verification test record.
1.5 GHz
100 kHz
b
5 MHz 10 kHz
1.5 GHz 5 MHz 55 MHz 100 kHz
1.5 GHz 55 MHz 1240 MHz 1 MHz
1.5 GHz 1240 MHz 1500 MHz 1 MHz
a. 75 RF Outputs: Omit this frequency range.
b. 75
RF Outputs: Set the start frequency to 1 MHz.
Table 2-102 1.5 GHz Tracking Generator Non-Harmonic Spurious Response
Worksheet
Analyzer
Center
Frequency
Microwave
Analyzer Start
Frequency
Microwave
Analyzer Stop
Frequency
Microwave
Analyzer
Resolution
Bandwidth
Amplitude of
Non-Harmonic
Response
(dBc)