Command Reference
Status Subsystem (:STATus)
Chapter 4190
Status Subsystem (:STATus)
:OPERation:CONDition
Supported All
:STATus:OPERation:CONDition?
This command returns the decimal sum of the bits for the registers that are set to one
and are part of the Standard Operation Status Group. For example, if a sweep is in
progress (bit 3), the value 8 is returned.
*RST N/A
Range 0–32767
Key Entry N/A
Remarks Refer to “Standard Operation Condition Register” on page 118 for more
information.
The data in this register is continuously updated and reflects current
conditions.
:OPERation:ENABle
Supported All
:STATus:OPERation:ENABle <value>
:STATus:OPERation:ENABle?
This command determines what bits in the Standard Operation Event Register will set
the Standard Operation Status Summary bit (bit 7) in the Status Byte Register.
The variable <value> is the sum of the decimal values of the bits that you want to
enable.
*RST N/A
Range 0–32767
Key Entry N/A
Remarks Refer to Standard Operation Status Group” on page 117 and
“Standard Operation Event Enable Register” on page 119 for more
information.