Command Reference
IEEE 488.2 Common Commands
Chapter 4 157
*OPC?
Supported All
*OPC?
The Operation Complete (OPC) query returns the ASCII character 1 when all pending
operations have finished.
*RST N/A
Range N/A
Key Entry N/A
Remarks N/A
*PSC
Supported All
*PSC ON|OFF|1|0
The Power-On Status Clear (PSC) command controls the automatic power-on clearing of
the Service Request Enable Register, the Standard Event Status Enable Register, and
device-specific event enable registers.
ON (1) This choice enables the power-on clearing of the listed registers.
OFF (0) This choice disables the clearing of the listed registers and they retain
their status when a power-on condition occurs.
*RST N/A
Key Entry N/A
Remarks The setting enabled by this command is not affected by signal
generator power-on, preset, or *RST.
Choices ON OFF 1 0