Agilent Technologies 1100 Series Instrukcja Obsługi Strona 63

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 272
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 62
63
Troubleshooting and Test Functions
Pressure Signal Missing
Pressure Signal Missing
The pressure signal from the damper is missing.
The pressure signal from the damper must be within a specific voltage range.
If the pressure signal is missing, the processor detects a voltage of
approximately -120mV across the damper connector.
Probable Causes Damper disconnected.
Defective damper.
Suggested Actions Ensure the damper is connected correctly to the LPM board.
Exchange the damper.