Troubleshooting and Test Functions
Shut-Down
Shut-Down
An external instrument has generated a shut-down signal on the remote line.
The quaternary pump continually monitors the remote input connectors for
status signals. A LOW signal input on pin 4 of the remote connector generates
the error message.
Probable Causes Leak detected in another module with a CAN connection to the system .
Leak detected in an external instrument with a remote connection to the
system.
Shut-down in an external instrument with a remote connection to the
system.
The degasser failed to generate sufficient vacuum for solvent degassing.
Suggested Actions Fix the leak in the external instrument before restarting the quaternary
pump.
Check external instruments for a shut-down condition.
Check the vacuum degasser for an error condition. Refer to the Reference
Manual for the Agilent 1100 Series vacuum degasser.