Timeout
The timeout threshold was exceeded.
Probable Causes The analysis was completed successfully, and the timeout function
switched off the quaternary pump as requested.
A not-ready condition was present during a sequence or multiple-injection
run for a period longer than the timeout threshold.
Suggested Actions Check the logbook for the occurrence and source of a not-ready condition.
Restart the analysis where required.