Repairing the Pump
Exchanging the Power Supply
Exchanging the Power Supply
When Required:
If defective
Tools required:
Screwdriver Pozidriv #1
Wrench 1/4 inch
Wrench 14 mm
Wrench 7 mm
Wrench 5 mm
Part required:
Power supply 0950-2528
Preparation for this procedure:
Switch off the pump at the main power
switch, remove all cables and tubings
1 Remove the pump from the stack, remove the
front cover, top cover and top foam section
(see Removing the Top Cover and Foam on
page 128).
2 Disconnect all capillaries, tubes and
connectors from the pump head and the
damper. Remove the MCDV cover.
3 Loosen the two Z-panel screws and remove
it.
Pump head
Damper
MCGV
Screws
Z-panel