Agilent Technologies DisplayPort Instrukcja Obsługi Strona 224

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 246
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 223
224 Agilent U7232A DisplayPort Electrical Performance Compliance Test Application
24 Calibrating the Infiniium Oscilloscope and Probe
manual comes together with the E2655A/B De- skew Kit, that came
with your oscilloscope.
SMA Probe Head Atten/Offset Calibration
1 Referring to Figure 90 below, perform the following steps:
a Connect the SMA to BNC adaptor to one of the SMA connectors of
the deskew fixture or SMA (f) to SMA (f) adapter.
b Connect the 50 ohm terminator to the center SMA connector of the
SMA probe head.
c Connect the other end of the deskew fixture or SMA (f) to SMA (f)
adapter to one of the SMA connectors of the SMA probe head.
d Connect the BNC connector of the SMA to BNC adaptor to the Aux
Out on the front panel of the Infiniium oscilloscope.
e Connect the positive (+) side of the InfiniiMax probe amp to the GPO
(SMP) connector of the SMA probe head whose path is connected to
Figure 89 De-skew Connection.
NOTE
Each probe is calibrated to the oscilloscope channel to which it is connected. Do not switch
probes between channels or other oscilloscopes, or it will be necessary to calibrate them
again. It is recommended that the probes are labeled with the channel on which they were
calibrated.
Przeglądanie stron 223
1 2 ... 219 220 221 222 223 224 225 226 227 228 229 ... 245 246

Komentarze do niniejszej Instrukcji

Brak uwag