184 Agilent U7232A DisplayPort Electrical Performance Compliance Test Application
21 Cable Eye Diagram Tests
Probing for Cable Eye Diagram Tests
When performing the cable eye diagram test, the DisplayPort Electrical
Performance Compliance Test Application will prompt you to make the
proper connections. Your DisplayPort test environment setup on the Set
Up tab must match the physical connection.
Figure 64 and Figure 65 show a physical connection for making
differential and single- ended connections.
.
Infiniium Oscilloscope
Lane on the test fixture
cables
DUT
Figure 64 Probing for Differential Tests - Cable Eye Diagram Tests (Single Connection with
DisplayPort Test Fixture)