Sink Non-ISI Jitter Tests 20
Agilent U7232A DisplayPort Electrical Performance Compliance Test Application 181
10 The following select file dialog box appears. Select your coefficient file
(*.equ) and click Open. The test will run based on the your user
defined coefficient file.
PASS Condition
UI is Unit Interval.
Figure 63 Selecting User Defined File
Table 24 Non-ISI Jitter at Internal and Compliance Points.
Receiver package pin Transmitter Connector (TP2)
High-bit Rate (2.7 Gb/s per lane)
A
p-p
0.339 UI 0.330 UI
Reduced-bit Rate (1.62 Gb/s per lane)
A
p-p
0.465 UI 0.442 UI