Agilent Technologies DisplayPort Instrukcja Obsługi Strona 113

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 246
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 112
Source Spread Spectrum Clocking (SSC) Differential Tests (Normative & Informative) 13
Agilent U7232A DisplayPort Electrical Performance Compliance Test Application 113
.
You can use any oscilloscope channel and connect it to any lane test point.
You select the channels used for testing lanes in the Set Up tab of the
DisplayPort Electrical Performance Compliance Test Application.
The data lanes and channels shown in the previous figures are just
examples. You can choose any desired data lane and channel that you
want.
For more information on the 1168A or 1169A probe amplifiers and
differential probe heads, see Chapter 25, “InfiniiMax Probing,” starting on
page 231.
Infiniium Oscilloscope
Data Channel
SMA to SMP cables
DUT
Lane on the W2641A test fixture
Figure 38 Differential Measurement Setup Using Two Single Ended Connections - SSC
Tests (A Minus B Configuration)
Przeglądanie stron 112
1 2 ... 108 109 110 111 112 113 114 115 116 117 118 ... 245 246

Komentarze do niniejszej Instrukcji

Brak uwag