9
Typical RMS Jitter
17
:
Standard
Carrier frequency SONET/SDH RMS jitter Unit Intervals Time (fs)
Data Rates bandwidth (µUI)
155 MHz 155 MB/s 100 Hz to 1.5 MHz 48 303
622 MHz 622 MB/s 1 kHz to 5 MHz 34 50
2.488 GHz 2488 MB/s 5 kHz to 15 MHz 65 25
9.953 GHz 9953 MB/s 20 kHz to 80 MHz 173 16
Option UNR
Carrier frequency SONET/SDH RMS jitter Unit Intervals Time (fs)
Data Rates bandwidth (µUI)
155 MHz 155 MB/s 100 Hz to 1.5 MHz 47 297
622 MHz 622 MB/s 1 kHz to 5 MHz 26 40
2.488 GHz 2488 MB/s 5 kHz to 15 MHz 66 25
9.953 GHz 9953 MB/s 20 kHz to 80 MHz 161 15
Maximum deviation N x 8 MHz
Resolution 0.1% of deviation or 1 Hz, whichever is greater
Deviation accuracy < ± 3.5% of FM deviation + 20 Hz
(1 kHz rate, deviations < N x 800 kHz)
Modulation frequency response
Path Rates (at 100 kHz deviation)
1 dB Bandwidth 3 dB Bandwidth, typical
FM 1 dc/20 Hz to 100 kHz dc/5 Hz to 10 MHz
FM 2 dc/20 Hz to 100 kHz dc/5 Hz to 1 MHz
dc FM
18
carrier offset ±0.1% of set deviation + (N x 8 Hz)
Distortion < 1% (1 kHz rate, deviations < N x 800 kHz)
Sensitivity ±1 V
peak
for indicated deviation
Paths FM1 and FM2 are summed internally for composite
modulation. Either path may be switched to any one of
the modulation sources: Ext1, Ext2, internal1, internal2.
The FM2 path is limited to a maximum rate of 1 MHz.
The FM2 path must be set to a deviation less than FM1.
Maximum deviation N x 80 radians (N x 8 radians in high-bandwidth mode)
Resolution 0.1% of set deviation
Deviation accuracy < ±5% of deviation + 0.01 radians
(1 kHz rate, normal BW mode)
Modulation frequency response
Mode Maximum Deviation Rates (3 dB BW)
Normal BW N x 80 rad dc – 100 kHz
High BW N x 8 rad dc – 1 MHz (typ)
Distortion
< 1 % (1 kHz rate, THD, dev < N x 80 rad, normal BW mode)
Sensitivity ±1 V
peak
for indicated deviation
Paths FM1 and FM2 are summed internally for composite
modulation. Either path may be switched to any one of
the modulation sources: Ext1, Ext2, internal1, internal2.
The FM2 path must be set to a deviation less than FM1.
Frequency modulation
Phase modulation
17 Calculated from phase noise performance in CW mode only at +0 dBm. For other frequencies, data rate, or
bandwidths, please contact your sales representative.
18 At the calibrated deviation and carrier frequency, within 5 °C of ambient temperature at time of user calibration.
Komentarze do niniejszej Instrukcji