3
Range
1
Option 520 250 kHz to 20 GHz
Option 540 250 kHz to 40 GHz
Resolution
CW 0.001 Hz
All Sweep modes 0.01 Hz
Accuracy Aging rate ± temperature effects
± line voltage effects
Switching speed
2
< 15 ms typical
Phase offset Adjustable in nominal 0.1° increments.
Frequency bands
Band Frequency range N #
1 250 kHz to 250 MHz 1/8
2 > 250 to 500 MHz 1/16
3 > 500 MHz to 1 GHz 1/8
4 > 1 to 2 GHz 1/4
5 > 2 to 3.2 GHz 1/2
6 > 3.2 to 10 GHz 1
7 > 10 to 20 GHz 2
8 > 20 to 40 GHz 4
Internal timebase reference oscillator
Standard Option UNR
Aging rate < ±1 x 10
–7
/year or < ±3 x10
–8
/year or
< ±4.5 x 10
–9
/day < ±2.5 x 10
–10
/day
after 45 days after 30 days
Temperature effects (typical) < ±5 x 10
–8
0 to 55° C < ±4.5 x 10
–9
0 to 55° C
Line voltage effects (typical) < ±2 x 10
–9
for < ±2 x 10
–10
for
+5% –10% change ±10% change
External reference frequency 1, 2, 2.5, 5, 10 MHz 10 MHz only
(within 1 ppm) (within 1 ppm)
Reference output
Frequency 10 MHz
Amplitude > +4 dBm typical into 50 Ω load
External reference input
Amplitude > –3 dBm
Opt UNR 5 dBm ±5 dB
3
Input impedance 50Ω, nominal
Specifications
Frequency
1 Useable to 100 kHz
2 To within 0.1 ppm of final frequency above 250 MHz or within 100 Hz below 250 MHz
3 To optimize phase noise 5 dBm ± 2 dB
Komentarze do niniejszej Instrukcji