812 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
32 Obsolete and Discontinued Commands
:MTESt:TRIGger:SOURce
(see page 860)
Command Syntax
:MTESt:TRIGger:SOURce <source>
<source> ::= CHANnel<n>
<n> ::= 1 to (# analog channels) in NR1 format
The :MTESt:TRIGger:SOURce command sets the channel to use as the
trigger.
Query Syntax
:MTESt:TRIGger:SOURce?
The :MTESt:TRIGger:SOURce? query returns the currently selected trigger
source.
Return Format
<source> ::= CHAN<n>
<n> ::= 1 to (# analog channels) in NR1 format
See Also "Introduction to :MTESt Commands" on page 387
NOTE
The :MTESt:TRIGger:SOURce command is obsolete and is provided for backward
compatibility to previous oscilloscopes. Use the trigger source commands (see page 641)
instead.