390 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
20 :MTESt Commands
:MTESt:ALL
(see page 860)
Command Syntax
:MTESt:ALL <on_off>
<on_off> ::= {{1 | ON} | {0 | OFF}}
The :MTESt:ALL command specifies the channel(s) that are included in the
mask test:
ON — All displayed analog channels are included in the mask test.
OFF — Just the selected source channel is included in the test.
Query Syntax
:MTESt:ENABle?
The :MTESt:ENABle? query returns the current setting.
Return Format
<on_off><NL>
<on_off> ::= {1 | 0}
See Also "Introduction to :MTESt Commands" on page 387