474 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
24 :SBUS<n> Commands
:SBUS<n>:CAN:TRIGger:PATTern:ID:MODE
(see page 860)
Command Syntax
:SBUS<n>:CAN:TRIGger:PATTern:ID:MODE <value>
<value> ::= {STANdard | EXTended}
The :SBUS<n>:CAN:TRIGger:PATTern:ID:MODE command sets the CAN
identifier mode. STANdard selects the standard 11-bit identifier. EXTended
selects the extended 29- bit identifier. The CAN identifier is set by the
:SBUS<n>:CAN:TRIGger:PATTern:ID command.
Query Syntax
:SBUS<n>:CAN:TRIGger:PATTern:ID:MODE?
The :SBUS<n>:CAN:TRIGger:PATTern:ID:MODE? query returns the current
setting of the CAN identifier mode.
Return Format
<value><NL>
<value> ::= {STAN | EXT}
Errors "- 241, Hardware missing" on page 823
See Also "Introduction to :TRIGger Commands" on page 641
":SBUS<n>:MODE" on page 457
":SBUS<n>:CAN:TRIGger:PATTern:DATA" on page 471
":SBUS<n>:CAN:TRIGger:PATTern:DATA:LENGth" on page 472
":SBUS<n>:CAN:TRIGger:PATTern:ID" on page 473