Agilent Technologies InfiniiVision 3000 DSO-X 3054A Instrukcja Obsługi Strona 555

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 970
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 554
:SBUS<n> Commands 24
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 555
:SBUS<n>:UART:TRIGger:TYPE
(see page 860)
Command Syntax
:SBUS<n>:UART:TRIGger:TYPE <value>
<value> ::= {RSTArt | RSTOp | RDATa | RD1 | RD0 | RDX | PARityerror
| TSTArt | TSTOp | TDATa | TD1 | TD0 | TDX}
The :SBUS<n>:UART:TRIGger:TYPE command selects the UART trigger type.
When one of the RD or TD types is selected, the
:SBUS<n>:UART:TRIGger:DATA and :SBUS<n>:UART:TRIGger:QUALifier
commands are used to specify the data value and comparison operator.
The RD1, RD0, RDX, TD1, TD0, and TDX types (for triggering on data and
alert bit values) are only valid when a 9- bit width has been selected.
Query Syntax
:SBUS<n>:UART:TRIGger:TYPE?
The :SBUS<n>:UART:TRIGger:TYPE? query returns the current UART
trigger data value.
Return Format
<value><NL>
<value> ::= {RSTA | RSTO | RDAT | RD1 | RD0 | RDX | PAR | TSTA |
TSTO | TDAT | TD1 | TD0 | TDX}
See Also "Introduction to :TRIGger Commands" on page 641
":TRIGger:MODE" on page 648
":SBUS<n>:UART:TRIGger:DATA" on page 552
":SBUS<n>:UART:TRIGger:QUALifier" on page 554
":SBUS<n>:UART:WIDTh" on page 556
Przeglądanie stron 554
1 2 ... 550 551 552 553 554 555 556 557 558 559 560 ... 969 970

Komentarze do niniejszej Instrukcji

Brak uwag