3-23
Verifying Performance
Agilent Technologies 8922M/S Specifications
Output RF Spectrum Measurement (Requires Option 006)
After one timeslot, 577µs, from an isolated Receiver Hop Trigger in the GSM
Frequency bands.
Dynamic Range (dB)
NOTE: When using Output RF Spectrum due to the Ramping measurement, the dynamic
range is decreased by 12 dB (due to Peak Hold).
Range
RF In/Out
6 to +41 dBm
Aux RF In
36 to +20 dBm
Input Levels for Optimum Dynamic Range
RF In/Out
+7, +17, +27, +37 dBm
Aux RF In
23, 13, 3, +7 dBm
Input Frequency Setting Error
±10 kHz,
Input Level Setting Error
±3 dB
Supplemental Characteristics:
Log Linearity
±0.4 dB
Amplitude Flatness
±1.0 dB
Amplitude Resolution
0.4 dB
Offset (kHz)
100 200 300 400 600 800 to 1800
Range (dB) 24 42 53 60 63 64