Making Measurements
If You Have Problems with a Measurement
Possible Solutions to
Sync Status Errors
Solution 1 - Trigger Timing
Check if the trigger is being received.
On the MEAS SYNC screen (press to gain access), check if the
correct burst type has been defined. On the Data Bits screen:
A The First Bit field on the Phase Freq:Data Bits screen displays the time
difference between when a trigger is being received and when the first bit of
a burst occurred (A). The time difference is only valid if FMErrCount is 0.
B Check the delay in the Trig Delay field (B).
Figure 2-18
MEAS SYNC
A
B