Output RF Spectrum, Trace View (Option 006 Only)
Output RF Spectrum, Trace View (Option 006 Only)
This view displays output RF spectrum power spectral density (at the Freq Offset
setting) versus time.
The Agilent 8922M/S uses a 3-pole synchronously tuned filter to make Output
RF Spectrum measurements rather than a 5-pole filter as specified in the GSM
recommendations. Refer to Section titled “Output RF Spectrum Measurements
Using a 3-Pole Synchronously Tuned Measurement Filter” in Chapter 2.
1. Amplitude This is a copy of the RF Analyzer Amplitude field. This is the amplitude to be
assumed at the selected input port.
Range RF IN/OUT: 27.9 to +41.0 dBm. AUX RF IN: 58.0 to +20.0
dBm.
See Also Screens: RF Generator/RF Analyzer (RF Analyzer)
2. Marker This field displays the level at the current marker position.
See Also Keys: ON/OFF, HI LIMIT, LO LIMIT, REF, AVG, Units