Agilent 81133A / 81134A Performance Verification
Agilent 81133A / 81134A Performance Verification, December 2003
8
4 Switch on the trigger output and set the counter to measure the
frequency at Channel 1
5 Check the frequency at the following settings:
Frequency Acceptable Measured Frequency TR Entry
15 MHz 14.99925 MHz to 15.00075 MHz 1-1
50 MHz 49.9975 MHz to 50.0025 MHz 1-2
100 MHz 99.995 MHz to 100.005 MHz 1-3
200 MHz 199.990 MHz to 200.010 MHz 1-4
6 Connect the 81133A/81134A Trigger Output to the Counter Channel 3
input.
7 Check the frequency at the following settings:
Frequency Acceptable Measured Frequency TR Entry
500 MHz 499.975 MHz to 500.025 MHz 1-5
1.0 GHz 999.950 MHz to 1.00005 GHz 1-6
2.0 GHZ 1.9999 GHz to 2.0001 GHz 1-7
3.0 GHz 2.99985 GHz to 3.00015 GHz 1-8
3.35 GHz 3.34983 GHz to 3.35016 GHz 1-9