Agilent 81133A / 81134A Performance Verification
Agilent 81133A / 81134A Performance Verification, December 2003
16
2 Set up the [Channel] screen as follows and switch on the channel
On the 81134A set up both channels and switch on only the channel that
is being tested.
3 Set up the Digitizing Oscilloscope Agilent 54121T:
Press [AUTOSCALE]
Select the Display menu and set the Number of Averages to 64
Select the Channel menu and set the Attenuation factor of channel 2 to
2
Set the VOLTS/DIV of channel 2 to 10 mV/div
Set OFFSET to 500 mV
Select the Timebase menu and set the TIME/DIV to 10 ps/div
4 Center the second positive edge of the signal
5 Press [MORE] and [HISTOGRAM]
Select the Window submenu and set:
Source is channel 2
Choose the Time Histogram
Press [WINDOW MARKER 1] and set it to 490 mV
Press [WINDOW MARKER 2] and set it to 500 mV