Agilent 81133A / 81134A Performance Verification
Agilent 81133A / 81134A Performance Verification, December 2003
29
On the 81134A set up both channels and switch on only the channel that
is being tested.
4 On the Oscilloscope:
Press [AUTOSCALE]
Select the Display menu and set the Number of Averages to 64
Select the Timebase menu and set TIME/DIV to 2 ns/div
Select the Delta V menu and switch on the V markers
Set the Preset Levels to Variable Levels and set the Variable Levels to
90 % and 110 %
Press Auto Level Set
Select the Timebase menu and set the TIME/DIV to 500 ps/div
Adjust the delay to center the positive edge
Select the Channels menu and set the channel 3 offset to 0.95 V
Set VOLTS/DIV to 200 mV/div
5 Check that the Overshoot of the positive edge is within the ± 10 % of
the amplitude limits and record the result in TR Entry 7-1