Agilent 81133A / 81134A Performance Verification
Agilent 81133A / 81134A Performance Verification, December 2003
27
6 Rise Time 20% - 80 %
On the Oscilloscope
Select the Timebase menu and set TIME/DIV to 5 ns/div
Select the Delta V menu and set the Preset Levels to 20% - 80%, press
Auto Level Set
Select the Timebase menu and set TIME/DIV to 20 ps/div
Adjust the delay until a positive edge is centered
Select the Delta t menu and press the Precise Edge Find key
7 Check that the Rise Time 20% - 80% is < 60 ps and record the measured
valued in TR entry 6-2
8 Fall Time 20% - 80 %
On the Oscilloscope
Select the Timebase menu and adjust the delay until a negative edge is
centered
Select the Delta t menu and set START ON EDGE = NEG 1 and STOP
ON EDGE = NEG 1, press the Precise Edge Find key
9 Check that the Fall Time 20% - 80% is < 60 ps and record the measured
valued in TR entry 6-3
10 Fall Time 10% - 90 %
On the Oscilloscope
Select the Timebase menu and set TIME/DIV to 5 ns/div
Select the Delta V menu and change the Preset Levels to 10% - 90%,
press Auto Level Set
Select the Timebase menu and set TIME/DIV to 20 ps/div
Select the Delta t menu and press the Precise Edge Find key
11 Check that the Fall Time 10% - 90% is < 90 ps and record the measured
valued in TR entry 6-4
For the 81134A repeat the test for channel 2.