Agilent Technologies 81134A Instrukcja Użytkownika Strona 17

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 35
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 16
Agilent 81133A / 81134A Performance Verification
Agilent 81133A / 81134A Performance Verification, December 2003
17
6 Select the Acquire submenu, set the Number of Samples to 1000 and
press [START ACQUIRING]
7 After the data for the time histogram has been acquired (#Samples =
100 %) select the Result submenu
8 Press [MEAN] and [SIGMA]. Record the values of sigma as the RMS
jitter.
9 Check the jitter at the following frequencies:
Frequency Acceptable Range TR Entry
100 MHZ < 4 ps 4-1
1 GHz < 4 ps 4-2
2 GHz < 4 ps 4-3
10 Set the Channel Frequency Divider of the 81133A/81134A to 2 and
check the clock jitter at 3 GHz. Record the result in TR Entry 4-4.
11 Data Jitter
Set up the 81133A/81134A as shown
Przeglądanie stron 16
1 2 ... 12 13 14 15 16 17 18 19 20 21 22 ... 34 35

Komentarze do niniejszej Instrukcji

Brak uwag