80 8000 Operator’s Manual
4 Operating the 8000
Once all values have been entered, press ESC. A confirmation screen
displays. Press Y to send the newly entered parameters to the Apparatus
3 / 7 and override the existing settings.
Main Menu Option 3 - Set Syringe Pump
This section is applicable only to systems including a syringe pump.
5 Com Port ID Select option 5, Com Port ID, to change the communication port identification
number. Your system was shipped with the communication port identification
number set to 01. You can set the identification number to any number between 1
and 99. The identification number is used when multiple units are connected in
series so that the controlling or master 8000 can start individual units.
Enter a communication port identification number and press ENTER. The Set
APP 3/7 Parameters screen displays.
6 Print Frequency The Apparatus 3 / 7 status prints at each sample time point. Select option 6, Print
Frequency, to set additional times to print the Apparatus 3 / 7 status between
sample time points. This option should be reserved for use when there is more
than 30 minutes between sample time points.
Enter a print frequency and press ENTER. The Set APP 3/7 Parameters screen
displays.
The print frequency timer resets after each sample time point.
Note: The status does not print if the specified print frequency time falls within
three minutes of the sample time point; the sample time point takes precedence.
Table 1 7 Set App 3/7 Parameters Options (Apparatus 3 / 7)
Option Function
Ensure the syringe pump is enabled prior to setting the syringe pump
parameters (see MENU > 0 > P under
“Hidden Key Functions” on
page 41
.