66 8000 Operator’s Manual
4 Operating the 8000
• If a shorter sample time point than the previous one is entered for a
given row, the row number will not increment. Times must increase
from one row to the next.
• To shorten a saved program, enter 000:00 for the first sample time
point you wish to discard. All successive time points are eliminated.
• Press ESC to save all new values and return to the Program Variables
screen.
Starting a 8000 Program
1 From the Ready screen, press START PROG to start a program or
modify a program already operating. The Enter Prog screen displays.
2 Enter a program number between 1 and 15 and press ENTER. The
Link Program Option screen displays.
Sample time points should be at least five minutes apart to allow
adequate time for priming and purging the sampling lines. If your
system configuration includes the syringe pump and filter changer,
ensure the sample time points are at least six minutes apart.
If using baskets or rotating cylinders, ensure the drive unit of the
dissolution apparatus is raised to its highest position before starting the
8000 program.
If using a dissolution apparatus with the 8000, you must enable external
control on the dissolution apparatus and log out.
***LINK PROGRAM OPTION***
1 RUN SINGLE PROG 2 LINK ANOTHER PROG
TIME: 01/01/12 15:02:30 ELAPSE: 000:00:04