Operating the 8000 4
8000 Operator’s Manual 99
Manual Calibration
To perform the calibration yourself, complete the following steps:
1 Remove the sample tray.
2 Ensure the sample cannulas of the dissolution apparatus are
sufficiently submerged in water.
3 Ensure the correct prime and purge times have been entered via the
Main Menu option 4 and 5, Manual Prime Time and Manual Purge
Time, respectively.
4 Press MENU.
5 Select option 7, Manual Operation.
6 Select option 4, Goto Rows.
7 Enter 4 to move the dispensing arm to row 4.
8 Remove the media rinse reservoir.
9 Press H to return to the home position.
10 Press ESC three times to return to the Ready screen.
11 From Ready screen, press CAL. The dispensing arm moves forward
to allow you to check the autocalibration block.
12 Remove the top portion of the autocalibration block and place it on
its side to prevent damage.
13 Place clean, empty 10 mL calibration tubes in the bottom portion of
the autocalibration block.
14 Select option 1, Manual. The dispensing arm moves over the
autocalibration block and primes the lines according to the manual
prime settings.
CAUTION
Do not use any acidic media to calibrate your 8000. Be sure to rinse
and dry the autocalibration block tips after completion.
NOTE
We suggest calibrating with tap water. Deionized water does not work
well with the autocalibration block which utilizes conductivity testing.