58 8000 Operator’s Manual
4 Operating the 8000
The following Manual Operations screen displays when your system
configuration includes the peristaltic pump:
The following Manual Operations screen displays when your system
configuration includes the syringe pump with or without the filter
changer:
All functions execute immediately without pressing ENTER. Press ESC
to return to the Main Menu.
Following is a description of the Manual Operations screen options:
***MANUAL OPERATIONS***
1 LIFT VALVES 2 LOWER VALVES
3 OPEN/CLOSE VALVES 4 GOTO ROWS
5 TURN ON (OFF) R-M PUMP
***MANUAL OPERATIONS***
1 LIFT VALVES 2 LOWER VALVES
3 OPEN/CLOSE VALVES 4 GOTO ROWS
5 TURN ON (OFF) R-M PUMP 6 SYRINGE/FILTER
Table 8 Manual Operations Options (Apparatus 1 / 2)
Option Function
1 Lift Valves Select option 1, Lift Valves, to raise the valve needles if lowered.
2 Lower Valves Select option 2, Lower Valves, to lower the valve needles if raised.
3 Open/Close Valves Select option 3, Open/Close Valves, to toggle the valves open and closed.
4 Goto Rows Select option 4, Goto Rows, to move the 8000 dispensing arm to a specific
row. Enter a row number. The dispensing arm moves to the indicated row and
stops.
If the moving alert is enabled, the alarm sounds when the dispensing arm
moves. To disable the alert, see MENU > 0 > B under “Hidden Key
Functions” on page 41.