Agilent Technologies E8267D PSG Instrukcja Użytkownika Strona 9

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 34
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 8
1. Specifications are typical for harmonics beyond specified frequency. Specifications are with Option 1EH low-pass filters below 2 GHz off, unless noted.
2. Below 250 MHz in ramp sweep mode (Option 007), Option 1EH filters are always off. Refer to harmonic specification with filters off.
Spectral purity
Harmonics
1
(dBc at +10 dBm or maximum specified output power, whichever is lower)
Frequency Standard
< 1 MHz –25 dBc (typ)
1 to < 10 MHz –25 dBc
10 to < 60 MHz –28 dBc
10 to < 60 MHz with Option 1EH filters on
2
–45 dBc
0.06 to 2 GHz –30 dBc
0.06 to 2 GHz with Option 1EH filters on
2
–55 dBc
> 2 to 20 GHz –55 dBc
> 20 to 44 GHz –45 dBc (typ)
Harmonics (measured)
Option 520 standard harmonics (measured)
Carrier frequency (GHz)
2nd harmonic (dBc)
-30
-40
-50
-60
-70
-80
0 2 4 6 8 10
Option 544 standard harmonics (measured)
Carrier frequency (GHz)
0
-20
-40
-60
-80
-100
-120
0 2 4 6 8 10 12 14 16 18 20 22
2nd harmonic (dBc)
2nd harmonic (dBc)
Standard vs. Option 1EH harmonics (measured)
Carrier frequency (MHz)
0
-20
-40
-60
-80
-100
0 500 1000 1500 2000
Standard
Option 1EH
9
Przeglądanie stron 8
1 2 3 4 5 6 7 8 9 10 11 12 13 14 ... 33 34

Komentarze do niniejszej Instrukcji

Brak uwag