Adjustable RF output limit
Frequency [GHz]
22
20
28
Power [dBm]
0
18
16
14
26
24
5101520
25 dBm
20 dBm
23 dBm
15 dBm
RF output limit (measured)
Resolution 0.01 dB
Temperature stability 0.01 dB/ °C (typ)
1
User flatness correction
Number of points 2 to 1601 points/table
Number of tables Up to 10,000, memory limited
Path loss Arbitrary, within attenuator range
Entry modes Remote power meter
2
, remote bus, manual (user edit/view)
Output impedance 50 Ω (nom)
SWR (internally leveled)
Option 520
250 kHz to 2 GHz < 1.4:1 (typ)
> 2 GHz to 20 GHz < 1.6:1 (typ)
Option 532 and 544
250 kHz to 1.2 GHz < 1.4:1 (typ)
> 1.2 GHz to 20 GHz < 1.6:1 (typ)
> 20 GHz < 1.8:1 (typ)
Leveling modes Internal leveling, external detector leveling, millimeter source module, ALC off
External detector leveling
Range –0.2 mV to –0.5 V (nom) (–36 dBm to +4 dBm using Agilent 33330D/E detector)
Bandwidth Selectable 0.1 to 100 kHz (nom) (note: not intended for pulsed operation)
Maximum reverse power 1/2 Watt, 0 V
DC
Adjustable RF output limit
Function Protects external devices by limiting maximum RF output; operates in all leveling modes
(internal, external, source module)
Range User-adjustable from +15 dBm to maximum output power
Accuracy
+15 to +25 dBm ±1 dB (typ)
> +25 dBm ±1.5 dB (typ)
Resolution 1 dB
Response time 30 μsec (measured)
Adjustment Can be locked to prevent accidental change
1. Options 532 and 544: 0.02 dB/°C (typ) above 2 GHz.
2. Compatible with Agilent EPM Series (E4418B and E4419B) power meters.