5
Table 1. System Dynamic Range
1
(Continued)
Description Specification (dB) at Test Port Typical (dB) at Test Port
Options
025, 125
Option
225
Option
425
Option
525
Options
025, 125
Option
225
Option
425
Option
525
Configurable Test Set and Extended Power Range
300 kHz to 3 MHz
2
92
3
-- -- -- -- -- -- --
3 MHz to 10 MHz 112 -- -- -- -- -- -- --
10 MHz to 45 MHz 121 -- -- -- -- 103 88 88
45 MHz to 70 MHz
4
121 101 90 90 -- -- -- --
70 MHz to 500
MHz
4
121 105 90 90 -- -- -- --
500 MHz to 2 GHz 121 110 110 110 -- -- -- --
2 GHz to 6 GHz 121 110 110 110 -- -- -- --
6 GHz to 8 GHz 120 110 110 110 -- -- -- --
8 GHz to 9 GHz 120 110 100 100 -- -- -- --
9 GHz to 10.5 GHz 116 110 100 100 --
10.5 GHz to 12.5
GHz
111 110 100 100 -- -- -- --
12.5 GHz to 13.5
GHz
108 108 100 100 -- -- -- --
13.5 GHz to 20 GHz -- 108 100 100 -- -- -- --
20 GHz to 31.25
GHz
-- -- 92 92 -- -- -- --
31.25 GHz to 40
GHz
-- -- 87 87 -- -- -- --
40 GHz to 50 GHz -- -- -- 75 -- -- -- --
1
The system dynamic range is calculated as the difference between the noise floor and the specified source maximum output power. The
effective dynamic range must take measurement uncertainties and interfering signals into account.
2
May be limited by crosstalk at certain frequencies below 3 MHz.
3
Value and frequency band changed July 2006.
4
May be degraded typically by 10 dB at particular frequencies (multiples of 5 MHz) below 500 MHz due to spurious receiver residuals.
Methods are available to regain the full dynamic range.
Receiver Dynamic Range technical specifications are not provided in this N5230A/C specs document.
Przeglądanie stron 8