7
N5230A/C Corrected System Performance with 3.5mm Connectors
Options 220/225, 420/425, 520/525: From 10 MHz to 45 MHz, performance is characterized as "typical". To generate these typical values,
please download our free Uncertainty Calculator from http://www.agilent.com/find/na_calculator
.
⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯
85052B Calibration Kit
⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯
Table 3. 85052B Calibration Kit
N5230A/C- Option 020/120 (Standard Test Set and Standard Power Range)
Applies to the N5230A/C Option 020/120 analyzers, 85052B (3.5mm) calibration kit, 85131F flexible test port cable set, and a full 2-port
calibration. Also applies to the following condition:
Environmental temperature 23° ±3 °C, with < 1 °C deviation from calibration temperature
Description Specification (dB)
300 kHz to
10 MHz
10 MHz to
2 GHz
2 GHz to
8 GHz
8 to
13.5 GHz
Directivity 48 48 44 44
Source Match 40 40 33 31
Load Match 48 48 44 44
Reflection Tracking
±0.003
+0.02/°C
±0.003
+0.02/°C
±0.003
+0.03/°C
±0.006
+0.03/°C
Transmission Tracking
±0.017
+0.02/°C
±0.015
+0.02/°C
±0.075
+0.03/°C
±0.131
+0.03/°C
Komentarze do niniejszej Instrukcji