Przeglądanie stron 34
31
⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯
N4693A Electronic Calibration Module
⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯
Table 11. N4693A Electronic Calibration Module
N5230A/C- Option 420 or 520 (Standard Test Set and Standard Power Range)
Applies to the N5230A/C Option420 or 520 analyzers, N4693A (2.4mm) Electronic Calibration Module, 85133F flexible test port cable set,
and a full 2-port calibration.
Also applies to the following condition:
Environmental temperature 23° ±3 °C, with < 1 °C deviation from calibration temperature
Typical (dB) Specification (dB) Description
10 to
200 MHz
45 MHz to
2 GHz
2 to
20 GHz
20 to
40 GHz
40 to
50 GHz
Directivity 32 55 49 43 41
Source Match 25 46 42 35 30
Load Match 24 43 41 37 36
Reflection Tracking
±0.050
+0.02/°C
±0.030
+0.02/°C
±0.040
+0.02/°C
±0.060
+0.02/°C
±0.080
+0.03/°C
Transmission Tracking
±0.10
+0.02/°C
±0.056
+0.02/°C
±0.079
+0.02/°C
±0.107
+0.02/°C
±0.130
+0.03/°C