52
Table 16. Test Port Output
1
(Continued)
Description Specifications Typicals
--
Options 020, 025, 120, 125, 220, 225, 420, 520, 425,
525
Non-Harmonic Spurious (at Nominal Output Power)
6
300 kHz to
10 MHz
--
10 MHz to
13.5 GHz
-- -50 dBc for offset frequency > 1 kHz
13.5 GHz to
20 GHz
--
20 GHz to
40 GHz
--
40 GHz to
50 GHz
-- -30 dBc for offset frequency > 1 kHz
Table 16. Test Port Output
1
(Continued)
Description Specifications Typical
--
Option
020,
025,
120,
125
Option
220,
225
Option
420, 520
Option
425, 525
Harmonics (2nd or 3rd) at Maximum Output Power
300 kHz to
10 MHz
-- -17 dBc -- -- --
10 MHz to
500 MHz
-- -17 dBc -22
dBc
-15 dBc -15 dBc
500 MHz to
1 GHz
-- -17 dBc -22
dBc
-15 dBc -15 dBc
1 GHz to
13.5 GHz
-- -20 dBc 22 dBc -20 dBc -20 dBc
13.5 GHz to
20 GHz
-- -- -22
dBc
-20 dBc -20 dBc
20 GHz to
40 GHz
-- -- -- -22 dBc -22 dBc
40 GHz to
50 GHz
-- -- -- -- -22 dBc
1
Performance specified on Port 1 only. Port 2 performance is a characteristic.
2
Power level linearity specified on Port 1 only. Port 2 performance is Typical. Test reference is at the nominal
power level.
3
ALC range starts at maximum leveled power and decreases in power level by the dB amount specified here.
4
Assumes no variation in time.
5
Assumes no variation in temperature.
6
Spurious signal levels are valid for a 10 MHz span centered on the carrier frequency. Spurious signals up to -15
dBc may exist outside the span. These signals do not affect the accuracy of the network analyzer
measurements.