548 Chapter 3
Performance Verification Test Records
Agilent E4405B Performance Verification Test Record
2.0 GHz 5600.0 MHz (5)________ –80 dBc ±.891 dB
2.0 GHz 6242.8 MHz (6)________ –80 dBc
±.891 dB
4.0 GHz 4042.8 MHz (7)________ –65 dBc
±.891 dB
4.0 GHz 4642.8 MHz (8)________ –65 dBc ±.891 dB
4.0 GHz 3742.9 MHz (9)________ –65 dBc
±.891 dB
4.0 GHz 2242.8 MHz (10)________ –80 dBc
±.891 dB
9.0 GHz 9042.8 MHz (11)________ –65 dBc
±.891 dB
9.0 GHz 9642.8 MHz (12)________ –65 dBc
±.891 dB
9.0 GHz 9342.8 MHz (13)________ –65 dBc
±.891 dB
9.0 GHz 4982.1 MHz (14)________ –80 dBc
±.891 dB
32. Spurious Responses
300 MHz TOI,
1 kHz RBW
12.5 dBm (1)________
±0.93 dB
300 MHz TOI, 30 Hz
RBW (Option 1DR only)
12.5 dBm (2)________
±0.93 dB
5 GHz TOI 11 dBm (3)________
±0.93 dB
8 GHz TOI 7.5 dBm (4)________
±0.93 dB
300 MHz SHI 35 dBm (5)________ ±0.41 dB
900 MHz SHI 45 dBm (6)________
±0.41 dB
1.55 GHz SHI 75 dBm (7)________
±0.41 dB
3.1 GHz SHI 90 dBm (8)________ ±0.41 dB
34. Gain Compression
Test Frequency
53 MHz (1)________ 1.0 dB
±0.123 dB
50.004 MHz
(Option 1DR only)
(2)________ 1.0 dB
±0.123 dB
Table 3-10 Agilent E4405B Performance Verification Test Record
Agilent Technologies
Model E4405B Report No. ___________
Serial No. ___________ Date ___________
Test Description Minimum
Results
Measured
Maximum
Measurement
Uncertainty