Agilent Technologies E1401-90493 Instrukcja Użytkownika Strona 451

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 642
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 450
Chapter 3 451
Performance Verification Test Records
Agilent E4402B Performance Verification Test Record
10 dB Step Attenuator ___________ ___________ ___________
Oscilloscope
(Option 1D6 only) ___________ ___________ ___________
Microwave Spectrum
Analyzer (Option 1DN only) ___________ ___________ ___________
Notes/comments: _____________________________________________________________
_____________________________________________________________
Table 3-4 Agilent E4402B Performance Verification Test Record
Agilent Technologies
Model E4402B Report No. ___________
Serial No. ___________ Date ___________
Test Description Minimum
Results
Measured
Maximum
Measurement
Uncertainty
1. 10 MHz Reference
Output Accuracy
(Non-Option 1D5 only)
Settability –5.0 Hz (1)________ 5.0 Hz
±293.3 µHz
2. 10 MHz High-Stability
Frequency Reference
Output Accuracy
(Option 1D5 only)
5 Minute Warm-Up Error –0.1 ppm (1)________ 0.1 ppm
±0.000072 ppm
15 Minute Warm-Up
Error
–0.01 ppm (2)________ 0.01 ppm
±0.000070 ppm
3. Frequency Readout and
Marker Frequency
Count Accuracy
Frequency Readout
Accuracy
Center Freq Span
1500 MHz 20 MHz 1499.83 MHz (1)________ 1500.17 MHz
±0 Hz
1500 MHz 10 MHz 1499.91 MHz (2)________ 1500.09 MHz ±0 Hz
1500 MHz 1 MHz 1499.991
MHz
(3)________ 1500.009
MHz
±0 Hz
Marker Count Accuracy
Table 3-3 Agilent E4402B Performance Verification Test Record