Chapter 3 465
Performance Verification Test Records
Agilent E4402B Performance Verification Test Record
Minimum Response –0.46 dB (2)________ ±0.19 dB
Peak-to-Peak Response (3)________ 0.92 dB
±0.19 dB
Option UKB
dc Coupled
Band 0
(100 Hz to 3.0 GHz)
Maximum Response (1)________ 0.50 dB
±0.19 dB
Minimum Response –0.50 dB (2)________
±0.19 dB
Peak-to-Peak Response (3)________ 1.0 dB
±0.19 dB
Non-Option UKB
Band 0A
(800 MHz to 1.0 GHz)
(serial number
US39441006 or greater)
Maximum Response (4)________ 0.46 dB
±0.11 dB
Minimum Response –0.46 dB (5)________
±0.11 dB
Peak-to-Peak Response (6)________ 0.92 dB
±0.11 dB
Non-Option UKB
Band 0B
(1.7 GHz to 2.0 GHz)
(serial number
US39441006 or greater)
Maximum Response (7)________ 0.46 dB
±0.11 dB
Minimum Response –0.46 dB (8)________ ±0.11 dB
Peak-to-Peak Response (9)________ 0.92 dB
±0.11 dB
Table 3-4 Agilent E4402B Performance Verification Test Record
Agilent Technologies
Model E4402B Report No. ___________
Serial No. ___________ Date ___________
Test Description Minimum
Results
Measured
Maximum
Measurement
Uncertainty