Agilent Technologies J-BERT N4903B Instrukcja Obsługi Strona 373

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 406
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 372
MEAS:JTOL:COMP(*):PAR:STANdard:DESCription[?]
MEAS:JTOL:COMP(*):PAR:STANdard:DESCription USER, <string>
MEAS:JTOL:COMP(*):PAR:STANdard:DESCription? [<standard enum>]
The command allows you to modify the description of the user-defined standard.
N O T E
The descriptions of predefined standards that have been provided by Agilent cannot
be modified.
The query returns the description of the standard specified by <standard enum>.
If no input parameter is given, the description of the currently selected standard is
returned.
The list of <standard enum> is returned by the MEAS:JTOL:COMP
(*):PAR:STANdard:NAMes? query.
MEAS:JTOL:COMP(*):PAR:MARGin
IAgilentN490xJComplianceMeasurement.Margin
MEAS:JTOL:COMP(*):PAR:MARGin <NR3>
The command sets the margin. The margin increases the jitter amplitudes of the
selected standard. A margin of 0.25, for example, has the effect that each amplitude
value is multiplied by 1.25.
MEAS:JTOL:COMP(*):PAR:RTIMe[?]
IAgilentN490xJComplianceMeasurement.DUTRelaxTime
MEAS:JTOL:COMP(*):PAR:RTIMe <NR3>
MEAS:JTOL:COMP(*):PAR:RTIMe?
The command sets the DUT relax time. This is the time before the measurement
proceeds to measure the BER after setting the jitter frequency and amplitude. The
next BER measurement starts after this time has elapsed. The unit is seconds.
The query returns the present value in seconds.
SCPI Command Reference 5
Agilent J-BERT N4903B High-Performance Serial BERT 373
Syntax
Description
IVI-COM Equivalent
Syntax
Description
IVI-COM Equivalent
Syntax
Description
Przeglądanie stron 372
1 2 ... 368 369 370 371 372 373 374 375 376 377 378 ... 405 406

Komentarze do niniejszej Instrukcji

Brak uwag