Agilent Technologies J-BERT N4903B Instrukcja Obsługi Strona 169

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 406
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 168
SOURce2:VOLTage[:LEVel][:IMMediate]:OFFSet[?]
IAgilentN490xPGClockVoltage.VOffset (IVI-compliant)
SOURce2:VOLTage[:LEVel][:IMMediate]:OFFSet <NR3>
SOURce2:VOLTage[:LEVel][:IMMediate]:OFFSet?
The command sets the offset value of the pattern generator's Clock Out port in
Volts. The query returns the offset value of the pattern generator's Clock Out port
in Volts.
SOURce2:VOLTage:LLEVel[?]
IAgilentN490xPGClockVoltage.LogicLevel (not IVI-compliant)
SOURce2:VOLTage:LLEVel <Family>
SOURce2:VOLTage:LLEVel?
<Family>: ECL | LVPecl | SCFL | LVDS | CML | CUSTom
N O T E
Selecting CUSTom has no effect.
ECL | LVPECL | SCFL | LVDS | CML | CUSTom
The command sets the output level appropriate for the specified logic family. The
query returns the currently used logic family.
N O T E
If any of the voltage parameters have been modified, CUSTom will be returned by
the query, even if the parameter has been set back to the default.
SOURce2:VOLTage:PROTection:RESet[?]
IAgilentN490xPGClockVoltage.ProtectionReset (not IVI-compliant)
SOURce2:VOLTage:PROTection:RESet[?]
SCPI Command Reference 5
Agilent J-BERT N4903B High-Performance Serial BERT 169
IVI-COM Equivalent
Syntax
Description
IVI-COM Equivalent
Syntax
Input Parameters
Return Range
Description
IVI-COM Equivalent
Syntax
Przeglądanie stron 168
1 2 ... 164 165 166 167 168 169 170 171 172 173 174 ... 405 406

Komentarze do niniejszej Instrukcji

Brak uwag