PLUGin[:J20]:GLOBal[:STATE][?]
IAgilentN490xJInterferenceEnable
PLUGin[:J20]:GLOBal[:STATE] ON | OFF | 1 | 0
PLUGin:GLOB?
Enables or disables the plug-in module.
When the module is enabled, the previous (present) setup is restored.
The query returns the present setting (0 | 1).
PLUGin[:J20]:SINTerference[:STATe][?]
IAgilentN490xJSinusoidalInterferenceEnable
PLUGin[:J20]:SINTerference[:STATE] ON | OFF | 1 | 0
PLUGin[:J20]:SINT?
Turns the generation of sinusoidal interference on or off.
The query returns the present setting (0 | 1).
PLUGin[:J20]:SINTerference:VOLTage[?]
IAgilentN490xJSinusoidalInterferenceAmplitude
PLUGin[:J20]:SINTerference:VOLTage[:LEVel][:IMMediate] [:AMPLitude] <NR3>
PLUGin[:J20]:SINT:VOLTage? MIN | MAX
Sets the sinusoidal interference voltage amplitude in V.
The query returns the present setting or the applicable min/max values.
PLUGin[:J20]:SINTerference:FREQuency[?]
IAgilentN490xJSinusoidalInterferenceModulationFrequency
5 SCPI Command Reference
314 Agilent J-BERT N4903B High-Performance Serial BERT
IVI-COM Equivalent
Syntax
Description
IVI-COM Equivalent
Syntax
Description