The command measures the DC level at the CDR input and sets the measured value
as CDR threshold.
The query returns the current DC level at the CDR input.
SENSe2:VOLTage:EDGe[?]
IAgilentN490xEDClockIn.ActiveEdge (IVI-compliant)
SENSe2:VOLTage:EDGe NEGative | POSitive
SENSe2:VOLTage:EDGe?
Sets the active edge of the clock input:
NEGative
The falling edge starts the period in which the input data is sampled.
POSitive
The rising edge starts the period in which the input data is sampled.
This command has restrictions for frequencies under 620Mbits/s. See for details.
See the User Guide (or online Help) for details.
SOURce7 Subsystem
SOURce7 Subsystem - Reference
SOURce7 represents the error detector's Trigger Out port.
SCPI Command Reference 5
Agilent J-BERT N4903B High-Performance Serial BERT 279
Description
IVI-COM Equivalent
Syntax
Description
Przeglądanie stron 278