Agilent Technologies E1445A Podręcznik Użytkownika Strona 71

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 539
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 70
Agilent E6432A Microwave Synthesizer Users Guide 1-57
Installation
Test 7. I/Q Functionality
E4430B Setup Summary E4430B Setup Details
Preset 1. Press the green Preset front panel key.
I/Q 2. Select the I/Q front panel key.
More (1 of 2) 3. Select the More (1 of 2) softkey.
I/Q Calibration 4. Select I/Q Calibration softkey.
Calibration Type Full 5. Select the Calibration Type Full softkey.
Execute Cal 6. Select the Execute Cal softkey.
7. Turn on I/Q Baseband Modulation.
E4430B Setup Summary E4430B Setup Details
Quick Set up 1. Press the Mode front panel key.
Mode = Real Time I/Q
Baseband
2. Select the Real Time I/Q Baseband softkey.
Custom Modulation = QPSK 3. Select the Custom softkey.
Symbol Rate = 2 MS/s 4. Select the Modulation Type softkey.
Filter = Root Nyquist 5. Select the Select (XXX) softkey where XXX is the
current modulation format selected.
Alpha = .5 6. Select the PSK (XXX) softkey where XXX is the
current modulation format selected.
Turn on Custom Modulation 7. Select the QPSK and OQPSK softkey.
8. Select the QPSK softkey.
9. Press the Return front panel key.
10. Select the Symbol Rate softkey.
11. Enter 2 MS/s using the numeric keypad and
softkey.
12. Press the Return front panel key.
13. Select the Filter softkey.
14. Select the Select (XXX) softkey, where XXX is the
current filter type selected.
15. Select the Root Nyquist softkey.
16. Select the Filter Alpha softkey.
17. Enter 0.5 via the numeric keypad.
18. Press the Return front panel key.
19. Select the Custom On softkey.
Przeglądanie stron 70
1 2 ... 66 67 68 69 70 71 72 73 74 75 76 ... 538 539

Komentarze do niniejszej Instrukcji

Brak uwag