2 If the message "Delay cal factors failed checksum–defaults
loaded
," appears on screen, go to "To perform the self-calibration, "
on page 4–25, perform steps 6 through 9, then continue to steps
below.
This message means that the horizontal delay calibration factors which were
stored in NVRAM along with a checksum have been corrupted. This
information is written into NVRAM when the horizontal delay calibration is
run. During powerup, the checksum was recomputed and compared against
the one saved when the horizontal delay calibration factors were saved. If
the checksums do not match, the error message is displayed.
a Cycle the power of the oscilloscope.
b If default horizontal delay calibration factors are loaded and error
messages remain, contact your Agilent service center for diagnosis
and repair.
3
If one of the following messages appears on screen:
"Results from last vertical calibration: Failed" or
"Results from last vertical calibration: Defaulted" or
"Results form last vertical calibration: Aborted",
go to "To perform the self-calibration," page 4–25 steps 1 through 5,
then continue to steps below.
The first message means that the most recent calibration of the vertical
subsystem failed.
The second message means that the default vertical factors have been loaded.
The third message means that the default vertical factors have been aborted
a Cycle the power of the oscilloscope.
b If default vertical calibration factors are loaded and error messages
remain, contact your Agilent service center for diagnosis and repair.
Service
Troubleshooting the Oscilloscope
4–38