6 Disconnect the ribbon cable from the display board, then check the
following signals on the system board.
Table 4-15 Signals from U609 (54615B and 54616B)
Signal Name Frequency Pulse width Voltage
U817 Pin 7 DE 19.72 kHz 38.0 µs 5.0 Vp-p
U817 Pin 24 Hsync 19.72 kHz 3.0 µs 5.0 Vp-p
J803 Pin 13 Vsync 60.00 Hz 253.5 µs 5.0 Vp-p
Signals from U609 (54616C)
Signal Name Frequency Pulse width Voltage
U827 Pin 7 DE 16.67 kHz 48.0 µs 5.0 Vp-p
U625 Pin 35 Hsync 16.67 kHz 2.90 µs 5.0 Vp-p
U625 Pin 20 Vsync 50.05 Hz 360.0 µs 5.0 Vp-p
If the signals are good, replace the display assembly. If not, replace the
system board.
7
Disconnect the ribbon cable from the display board.
8 Go to "To check the Low Voltage Power Supply," on page 4-40.
If the voltages are within the test limits, replace the display assembly. If not,
do the steps below.
a Disconnect the power cord.
b Disconnect the ribbon cable from the power supply.
c Connect the dummy load to the power supply connector.
d Connect the power cord, then measure the power supply voltages
again (see new tolerances below).
If the voltages are now within the test limits, replace the system
board. If not, replace the power supply.
Service
Troubleshooting the Oscilloscope
4–35