Test Re cords Performance Tests
Agilent 81600B Tunable Laser Source Family, Fourth Edition 187
Agilent 81600B #150 Performance Test Page 11 of 14
Model Agilent 81600B #150 Tunable Laser Report No. ________ Date_______
Power Stability
1
Power Stability = Maximum Deviation – Minimum Deviation
Signal-to-Source Spontaneous Emission - Output 2, High Power
Measurement Uncertainty: ± 0.20 dB
Low SSE Output 1 High Power Output 2
Att = 0 dB
Maximum Deviation dB dB
Minimum Deviation dB dB
Power Stability
1
dB dB
Upper Test Limit 0.02 dBpp 0.02 dBpp
Measurement Uncertainty ±0.005 dB ±0.005 dB
Wavelength Output Power Results Lower Test Limit
1450.000 nm - 1.00 dBm dB 37 dB
1460.000 nm - 1.00 dBm dB 37 dB
1470.000 nm - 1.00 dBm dB 37 dB
1480.000 nm + 4.00 dBm dB 42 dB
1490.000 nm + 4.00 dBm dB 42 dB
1500.000 nm + 4.00 dBm dB 42 dB
1510.000 nm + 4.00 dBm dB 42 dB
1520.000 nm + 5.00 dBm dB 45 dB
1530.000 nm + 5.00 dBm dB 45 dB
1540.000 nm + 5.00 dBm dB 45 dB
1550.000 nm + 5.00 dBm dB 45 dB
1560.000 nm + 5.00 dBm dB 45 dB
1570.000 nm + 5.00 dBm dB 45 dB
1580.000 nm - 4.00 dBm dB 42 dB
1590.000 nm - 1.00 dBm dB 37 dB