Performance Tests Power Tests
102 Agilent 81600B Tunable Laser Source Family, Fourth Edition
Test Procedure - High Power output, no attenuation
This test procedure is applicable to the:
It measures the power flatness of the module at the outputs indicated.
1 Set up the equipment as shown in Figure 30:
Figure 30 Test Setup for Power Flatness Tests - High Power output, without using
attenuator
2 Move to the TLS channel of the Agilent 8164A/B Lightwave
Measurement System and set the [Menu] parameters to:
Agilent 81600B #200 (Output 2, High Power), built-in attenuator not used.
Agilent 81600B #160 (Output 2, High Power), built-in attenuator not used.
Agilent 81600B #150 (Output 2, High Power), built-in attenuator not used.
Agilent 81600B #140 (Output 2, High Power), built-in attenuator not used.
Agilent 81600B #130 (Output 2, High Power), built-in attenuator not used.
Agilent 81600B #142 #003, built-in attenuator not used.
Agilent 81600B #142
Agilent 81600B #132
Tunable Laser Channel [Menu] Parameters Values
<Wavelength Mode><λ>
<Source State><Off>
<Power Unit><dBm>
<Power Mode>
a
a
For the 81600B #200, #160, #150, #140, and #142 #003 (that is, where a built-in
attenuator is fitted).
<Manual Att>
Modulation: <mod src><Off>