324 Chapter 9
O:\Manuals\E6380A_Progguid\Book\Errormsg.fm
Error Messages
Numbered HP-IB Error Descriptions
Example 9-125 Error -1316
Incomplete data received on RECC for word [1,2,3].
Indicates that the decoded data received on the reverse control channel
did not contain the proper number of bits in word 1 and/or word 2
and/or word 3. The raw decoded data is displayed in hexadecimal
format in the top right-hand portion of the CALL CONTROL screen. Raw
decoded data is only displayed when the CALL CONTROL screen Display
field is set to Data.
Example 9-126 Error -1317
Incomplete data received on RVC for word [1,2,3].
Indicates that the decoded data received on the reverse voice channel
did not contain the proper number of bits in word 1 and/or word 2
and/or word 3. The raw decoded data is displayed in hexadecimal
format in the top right-hand portion of the CALL CONTROL screen. Raw
decoded data is only displayed when the CALL CONTROL screen Display
field is set to Data.